Availability of Thickness Estimation of the Subcutaneous Fat by Using the Near-infrared Ray Measuring Device

Shigeaki Masuda1*, Tetsuya Tanioka2, Yuko Yasuhara2, Asumi Atsuta3, Hirokazu Ito2, Kazushi Motoki4, Kensaku Takase5, Beth King6 and Rozzano Locsin2

1Faculty of Nursing, School of Medicine, Nara Medical University, Nara, Japan
2Department of Nursing, Institute of Biomedical Sciences, Tokushima University,Tokushima, Japan
3Departmentof Nursing, Institute of Biomedical Sciences, Tokushima University Graduate School, Tokushima, Japan
4Tokushima Prefectural Kaifu Hospital, Tokushima, Japan
5Tokushima Prefectural Central Hospital, Tokushima, Japan
6Christine E. Lynn College of Nursing, Florida Atlantic University, 777 Glades Rd, Boca Raton, FL 33431 USA
Shigeaki Masuda, Faculty of Nursing, School of Medicine, Nara Medical University, 840 Shijocho, Kashihara, Nara Prefecture 634-0813 Japan; E-mail: masuda@naramed-u.ac.jp
16 September 2016; 24 November 2016; 26 November 2016
Masuda S, Tanioka T, Yasuhara Y, Atsuta A, Ito H, et al. (2016) Availability of Thickness Estimation of the Subcutaneous Fat by Using the Nearinfrared Ray Measuring Device. Int J Nurs Clin Pract 3: 210. doi: https://doi.org/10.15344/2394-4978/2016/210

Abstract

Gluteal intramuscular injection requires accurate needle insertion into the specific muscle area for accurate, safe, and efficacious delivery of medication for long-acting anti-psychiatric drugs. With ultrasonography, accurate evaluation of the "Distance from the Epidermis to the Under-Fascia (DEUF)" can be determined. However, a nurse in the clinical settings, ultrasonographic measurements during intramuscular injection is impractical and difficult. The purpose of this study was to estimate the thickness of subcutaneous fat by using a near-infrared ray measuring device.This study was approved by the Ethics Committee of Tokushima University Hospital. There were 39 subjects (21 men and 18 women) aged 20 to 60 years (Mean ± SD: 44.97 ± 15.20 years old). Bilateral DEUF at Ventrogluteal and Dorsogluteal sites (AVIUS: Hitachi-Aloka Medical Co.) were evaluated on B-mode scan as the gold standard. At the same sites, the thickness of the subcutaneous fatty tissue was measured using the PoccoStick (HORIC: smallsize instrument measuring the thickness of the subcutaneous fat with a near-infrared ray device). The right Ventrogluteal site thickness of the fatty tissue was significantly overestimated by PoccoStick (15.18 ± 8.77 mm) measurements, as compared to the ultrasonography (12.42 ± 5.26 mm) (p<0.01) results. The left-side Ventrogluteal site and both the right and left Dorsogluteal sites measurements did not show significant differences. This illustrated and supported the credible use of inexpensive devices such as the PoccoStick in estimating the DEUF. The development of a more highly accurate small-sized instrument is recommended.